Verification of Quantitative Measurement of Urinary Creatinine Enzymatic Method Kit by DIRUI CS-T240 Clinical Chemistry Analyser
DOI:
https://doi.org/10.31344/ijhhs.v0i0.171Keywords:
Creatinine, Method verification, DIRUI CS-T240, Enzymatic, Jaffe methodAbstract
Introduction: All urine samples sent for the diagnosis of inborn errors of metabolism (IEM) disorders testing will be first analysed for their creatinine levels. The urine creatinine analysis is applied either for calculation of urine amount needed in other tests or for quantitation of analytes per mol creatinine. Jaffe kinetic-end point method has been the preferred method. We aim to verify the performance of DIRUI enzymatic method reagent kit on DIRUI CS-T240 analyser and compare with Randox Jaffe kinetic-end point method reagent kit on same platform.
Method: Precision study, linearity and method comparisons were carried out using three levels of internal quality control (IQC) samples and the measured results were compared against manufacturer’s claims, medical decision limit and biological variation (BV).
Results: The within run coefficient variation (CV) were 3.57%, 2.61% and 1.35% for IQC Level 1, 2 and 3 respectively. The total precision and accuracy were 9.59% and 0.27% for Level 1, 3.17% and 0.57% for Level 2, while 2.03% and 3.63% for Level 3. Whereas, the verifica t ion value for CV based on analytical performance specifications (APS) BV desirable for CV and accuracy were 18.2% and 12.2% respectively. These values were compared, and the precision study was acceptable within desirable specification. Linearity met the manufacturer’s claim and range from 0.32 to 21.2mmol/L. The method comparison gave good correlation, r2 of 0.9975 and regression equation, y=0.9852x-0.022 with medical decision limit, 0.53mmol/L. The assessment of acceptability of this new method was acceptable as calculated total error (TEc) < total allowable error (TEa) (APS BV Desirable for TEa = 42.1%).
Conclusion: The analytical verification study that we had carried out fulfilled the BV desirable in precision, accuracy and TEa. Therefore, the performances of enzymatic method kit on DIRUI CS-T240 analyser is verified to be fit-for-purpose and can replace the existing method.
International Journal of Human and Health Sciences Supplementary Issue: 2019 Page: 53
Downloads
Published
How to Cite
Issue
Section
License
Authors who publish in the International Journal of Human and Health Sciences agree to the following terms that:
- Authors retain copyright and grant International Journal of Human and Health Sciences the right of first publication of the work.

Articles in International Journal of Human and Health Sciences are licensed under a Creative Commons Attribution 4.0 International License CC BY-4.0.This license permits use, distribution and reproduction in any medium, provided the original work is properly cited.- Authors are able to enter into separate, additional contractual arrangements for the distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted to post their work online (e.g., in institutional repositories or on their website) as it can lead to productive exchanges, as well as greater citation of published work.